HEAL DSpace

New aspects on the role of blood pressure and arterial stiffness in mechanical assistance by intra-aortic balloon pump: In-vitro data and their application in clinical practice

Αποθετήριο DSpace/Manakin

Εμφάνιση απλής εγγραφής

dc.contributor.author Papaioannou, TG en
dc.contributor.author Mathioulakis, DS en
dc.contributor.author Stamatelopoulos, KS en
dc.contributor.author Gialafos, EJ en
dc.contributor.author Lekakis, JP en
dc.contributor.author Nanas, J en
dc.contributor.author Stamatelopoulos, SF en
dc.contributor.author Tsangaris, SG en
dc.date.accessioned 2014-03-01T01:21:07Z
dc.date.available 2014-03-01T01:21:07Z
dc.date.issued 2004 en
dc.identifier.issn 0160-564X en
dc.identifier.uri https://dspace.lib.ntua.gr/xmlui/handle/123456789/16073
dc.subject intra-aortic balloon en
dc.subject counterpulsation en
dc.subject arterial stiffness en
dc.subject mock circulation en
dc.subject.classification Engineering, Biomedical en
dc.subject.classification Transplantation en
dc.subject.other AORTIC COUNTERPULSATION en
dc.subject.other CARDIOGENIC-SHOCK en
dc.subject.other HEART-FAILURE en
dc.subject.other CIRCULATION en
dc.subject.other MODEL en
dc.subject.other PERFORMANCE en
dc.subject.other AGE en
dc.title New aspects on the role of blood pressure and arterial stiffness in mechanical assistance by intra-aortic balloon pump: In-vitro data and their application in clinical practice en
heal.type journalArticle en
heal.identifier.primary 10.1111/j.1525-1594.2004.00080.x en
heal.identifier.secondary http://dx.doi.org/10.1111/j.1525-1594.2004.00080.x en
heal.language English en
heal.publicationDate 2004 en
heal.abstract Despite the well-known beneficial effects of the intra-aortic balloon pump (IABP) generally, there are still some clinical conditions accompanied by IABP ineffectiveness. The aim of this study was the investigation of the independent effects of arterial stiffness and blood pressure on acute IABP effectiveness. For this purpose, a mock circulatory system and 20 patients with cardiogenic shock due to acute myocardial infarction, were employed. It was shown that IABP acute efficiency was determined primarily by arterial compliance (AC) rather than blood pressure alone. IABP induced low hemodynamic effects in patients with systolic blood pressure > 80 mm Hg but with increased AC, whereas IABP resulted in greater hemodynamic effectiveness in cases with systolic pressure < 70 mm Hg but lower AC. The present study provides evidence concerning the hemodynamic conditions, which might lead to optimization of IABP or to the prediction of its acute hemodynamic performance, based on both measurements of AC and blood pressure. en
heal.publisher BLACKWELL PUBLISHING INC en
heal.journalName ARTIFICIAL ORGANS en
dc.identifier.doi 10.1111/j.1525-1594.2004.00080.x en
dc.identifier.isi ISI:000222839000004 en
dc.identifier.volume 28 en
dc.identifier.issue 8 en
dc.identifier.spage 717 en
dc.identifier.epage 727 en