HEAL DSpace

Comparison of CsI(Tl) and CsI(Na) partially slotted crystals for high-resolution SPELT imaging

Αποθετήριο DSpace/Manakin

Εμφάνιση απλής εγγραφής

dc.contributor.author Giokaris, N en
dc.contributor.author Loudos, G en
dc.contributor.author Maintas, D en
dc.contributor.author Karabarbounis, A en
dc.contributor.author Lembesi, M en
dc.contributor.author Spanoudaki, V en
dc.contributor.author Stiliaris, E en
dc.contributor.author Boukis, S en
dc.contributor.author Sakellios, N en
dc.contributor.author Karakatsanis, N en
dc.contributor.author Gektin, A en
dc.contributor.author Boyarintsev, A en
dc.contributor.author Pedash, V en
dc.contributor.author Gayshan, V en
dc.date.accessioned 2014-03-01T01:23:43Z
dc.date.available 2014-03-01T01:23:43Z
dc.date.issued 2006 en
dc.identifier.issn 0168-9002 en
dc.identifier.uri https://dspace.lib.ntua.gr/xmlui/handle/123456789/17111
dc.subject scintillators en
dc.subject partially slotted crystals en
dc.subject SPELT en
dc.subject PSPMTs en
dc.subject high-resolution en
dc.subject.classification Instruments & Instrumentation en
dc.subject.classification Nuclear Science & Technology en
dc.subject.classification Physics, Particles & Fields en
dc.subject.classification Spectroscopy en
dc.subject.other GAMMA-CAMERA en
dc.subject.other DETECTOR en
dc.subject.other SCINTIMAMMOGRAPHY en
dc.subject.other PSPMT en
dc.subject.other SCINTILLATOR en
dc.subject.other ARRAY en
dc.subject.other SPECT en
dc.title Comparison of CsI(Tl) and CsI(Na) partially slotted crystals for high-resolution SPELT imaging en
heal.type journalArticle en
heal.identifier.primary 10.1016/j.nima.2006.08.014 en
heal.identifier.secondary http://dx.doi.org/10.1016/j.nima.2006.08.014 en
heal.language English en
heal.publicationDate 2006 en
heal.abstract Dedicated systems based on Position Sensitive Photomultiplier Tubes (PSPMTs) coupled to scintillators, have been used over the past years for the construction of compact systems, suitable for applications such as small animal imaging and small organs imaging. Most of the proposed systems are based on fully pixelized scintillators. Previous studies have shown that partially slotted scintillators offer a good compromise between cost, energy resolution and spatial resolution. In this work, the performance of two sets of Csl(TI) and Csl(Na) partially slotted crystals is compared. Initial results show that CsI(TI) scintillators are more suitable for gamma-ray detection, since their performance in terms of sensitivity, spatial and energy resolution is superior than that of Csl(Na). (c) 2006 Elsevier B.V. All rights reserved. en
heal.publisher ELSEVIER SCIENCE BV en
heal.journalName NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT en
dc.identifier.doi 10.1016/j.nima.2006.08.014 en
dc.identifier.isi ISI:000243241300009 en
dc.identifier.volume 569 en
dc.identifier.issue 2 en
dc.identifier.spage 185 en
dc.identifier.epage 187 en