HEAL DSpace

Aspects of GEANT4 Monte-Carlo calculations of the BC501A neutron detector

Αποθετήριο DSpace/Manakin

Εμφάνιση απλής εγγραφής

dc.contributor.author Patronis, N en
dc.contributor.author Kokkoris, M en
dc.contributor.author Giantsoudi, D en
dc.contributor.author Perdikakis, G en
dc.contributor.author Papadopoulos, CT en
dc.contributor.author Vlastou, R en
dc.date.accessioned 2014-03-01T01:25:56Z
dc.date.available 2014-03-01T01:25:56Z
dc.date.issued 2007 en
dc.identifier.issn 0168-9002 en
dc.identifier.uri https://dspace.lib.ntua.gr/xmlui/handle/123456789/17833
dc.subject GEANT4 en
dc.subject Monte-Carlo simulations en
dc.subject Neutron spectroscopy en
dc.subject Scintillation detectors en
dc.subject.classification Instruments & Instrumentation en
dc.subject.classification Nuclear Science & Technology en
dc.subject.classification Physics, Particles & Fields en
dc.subject.classification Spectroscopy en
dc.subject.other Computer simulation en
dc.subject.other Monte Carlo methods en
dc.subject.other Neutron spectrometers en
dc.subject.other Protons en
dc.subject.other Scintillation counters en
dc.subject.other GEANT4 en
dc.subject.other Knockout protons en
dc.subject.other Monte Carlo simulations en
dc.subject.other Neutron energies en
dc.subject.other Neutron detectors en
dc.title Aspects of GEANT4 Monte-Carlo calculations of the BC501A neutron detector en
heal.type journalArticle en
heal.identifier.primary 10.1016/j.nima.2007.05.151 en
heal.identifier.secondary http://dx.doi.org/10.1016/j.nima.2007.05.151 en
heal.language English en
heal.publicationDate 2007 en
heal.abstract The response function of a commercial 2 in. x 2 in. BC501A neutron detector has been studied through the general-purpose Monte-Carlo (MC) simulation toolkit GEANT4. For neutron energies below 8 MeV, where the dominant detection mechanism originates from the knockout protons, the results of MC calculations reproduce remarkably well the experimental pulse-height spectra. This is not the case, however, at higher neutron energies, where the contribution of heavier charged particles to the detection mechanism becomes significant. The calculated detector efficiency is also compared at three indicative energies to the experimental one via the neutron activation technique. (c) 2007 Elsevier B.V. All rights reserved. en
heal.publisher ELSEVIER SCIENCE BV en
heal.journalName Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment en
dc.identifier.doi 10.1016/j.nima.2007.05.151 en
dc.identifier.isi ISI:000248784500038 en