dc.contributor.author |
Loudos, G |
en |
dc.contributor.author |
Majewski, S |
en |
dc.contributor.author |
Wojcik, R |
en |
dc.contributor.author |
Weisenberger, A |
en |
dc.contributor.author |
Sakellios, N |
en |
dc.contributor.author |
Nikita, K |
en |
dc.contributor.author |
Uzunoglu, N |
en |
dc.contributor.author |
Bouziotis, P |
en |
dc.contributor.author |
Varvarigou, A |
en |
dc.date.accessioned |
2014-03-01T01:26:54Z |
|
dc.date.available |
2014-03-01T01:26:54Z |
|
dc.date.issued |
2007 |
en |
dc.identifier.issn |
0168-9002 |
en |
dc.identifier.uri |
https://dspace.lib.ntua.gr/xmlui/handle/123456789/18272 |
|
dc.subject |
Dynamic studies |
en |
dc.subject |
Mouse-sized camera |
en |
dc.subject |
PSPMTs |
en |
dc.subject |
Radiopharmaceuticals evaluation |
en |
dc.subject |
Small animal imaging |
en |
dc.subject.classification |
Instruments & Instrumentation |
en |
dc.subject.classification |
Nuclear Science & Technology |
en |
dc.subject.classification |
Physics, Particles & Fields |
en |
dc.subject.classification |
Spectroscopy |
en |
dc.subject.other |
Drug products |
en |
dc.subject.other |
Imaging systems |
en |
dc.subject.other |
Optical resolving power |
en |
dc.subject.other |
Photodetectors |
en |
dc.subject.other |
Radiotherapy |
en |
dc.subject.other |
Dynamic studies |
en |
dc.subject.other |
Mouse-sized camera |
en |
dc.subject.other |
Radiopharmaceuticals evaluation |
en |
dc.subject.other |
Small animal imaging |
en |
dc.subject.other |
Cameras |
en |
dc.title |
Performance evaluation of a mouse-sized camera for dynamic studies in small animals |
en |
heal.type |
journalArticle |
en |
heal.identifier.primary |
10.1016/j.nima.2006.10.026 |
en |
heal.identifier.secondary |
http://dx.doi.org/10.1016/j.nima.2006.10.026 |
en |
heal.language |
English |
en |
heal.publicationDate |
2007 |
en |
heal.abstract |
A mouse sized camera has been built in terms of collaboration between the presenting institutions. The system is used for the performance of dynamic studies in small animals, in order to evaluate novel radiopharmaceuticals. The active area of the detector is approximately 48 x 96 mm allowing depiction of the entire mouse in a single view. The system is based on two flat-panel Hamamatsu H8500 position sensitive photomultiplier tubes (PSPMT), a pixellated NaI(Tl) scintillator and a copper-beryllium (CuBe) parallel-hole collimator. In this work, the evaluation results of the system are presented, using phantoms and small animals injected with conventional radiophrmaceuticals. Average resolution was similar to 1.6 mm on the collimator surface and increased to similar to 4.1 mm in 12cm distance from the detector. The average energy resolution was measured and found to be similar to 15.6% for Tc-99m. Results from imaging thin capillaries demonstrated system's high resolution and sensitivity in activity variations was shown. Initial dynamic studies have been carried out in small animals injected with Tc-99m-DTPA and Tc-99m-MDP. The results show system's ability to perform kinetic imaging in small animals. (c) 2006 Published by Elsevier B.V. |
en |
heal.publisher |
ELSEVIER SCIENCE BV |
en |
heal.journalName |
Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment |
en |
dc.identifier.doi |
10.1016/j.nima.2006.10.026 |
en |
dc.identifier.isi |
ISI:000244599100012 |
en |
dc.identifier.volume |
571 |
en |
dc.identifier.issue |
1-2 SPEC. ISS. |
en |
dc.identifier.spage |
48 |
en |
dc.identifier.epage |
51 |
en |