Investigation on the distal screw of a trochanteric intramedullary implant (Fi-nail) using a simplified finite element model

Αποθετήριο DSpace/Manakin

Εμφάνιση απλής εγγραφής

dc.contributor.author Efstathopoulos, N en
dc.contributor.author Nikolaou, VS en
dc.contributor.author Xypnitos, FN en
dc.contributor.author Korres, D en
dc.contributor.author Lazarettos, I en
dc.contributor.author Panousis, K en
dc.contributor.author Kasselouris, EN en
dc.contributor.author Venetsanos, DT en
dc.contributor.author Provatidis, CG en
dc.date.accessioned 2014-03-01T01:33:40Z
dc.date.available 2014-03-01T01:33:40Z
dc.date.issued 2010 en
dc.identifier.issn 0020-1383 en
dc.identifier.uri https://dspace.lib.ntua.gr/xmlui/handle/123456789/20526
dc.subject Intramedullary nail en
dc.subject Distal screw en
dc.subject Finitive element method en
dc.subject Fi-nail en
dc.subject Biomechanical behaviour en
dc.subject.classification Critical Care Medicine en
dc.subject.classification Emergency Medicine en
dc.subject.classification Orthopedics en
dc.subject.classification Surgery en
dc.subject.other DYNAMIC HIP SCREW en
dc.subject.other PROXIMAL FEMORAL NAIL en
dc.subject.other GAMMA-NAIL en
dc.subject.other INTERTROCHANTERIC FRACTURES en
dc.subject.other PERITROCHANTERIC FRACTURES en
dc.subject.other SUBTROCHANTERIC FRACTURES en
dc.subject.other INTERNAL-FIXATION en
dc.subject.other CONTROLLED TRIAL en
dc.subject.other FEMUR en
dc.subject.other PLATE en
dc.title Investigation on the distal screw of a trochanteric intramedullary implant (Fi-nail) using a simplified finite element model en
heal.type journalArticle en
heal.identifier.primary 10.1016/j.injury.2009.09.006 en
heal.identifier.secondary http://dx.doi.org/10.1016/j.injury.2009.09.006 en
heal.language English en
heal.publicationDate 2010 en
heal.abstract Numerous studies have been published conceit-ling the characteristics and the behaviour of the intramedullary devices in the treatment of the intertrochanteric hip fractures However, there is still room for further exploration and exploitation concerning the implant behaviour with respect to the par is of the implant assembly (nail, lag screw and distal screw) Towards this direction, the present paper armed at revealing the effect of the position of the distal screw on the mechanical behaviour of the fixation device For this purpose, a simplified model was developed and analysed with the finite element method In total, five different locations for the distal screw were examined In all cases, the bone was fixed at its distal end while the external load was applied at the tip of the lag screw towards the hip and in the form of orthonormal force components applied individually The results of the FE analyses were illustrated in appropriately formed plots revealing the sensitivity of the behaviour of the implant with respect to the location of the distal screw The main conclusion derived from the present investigation was that moving the distal screw apically decreases the stresses on the distal screw but increases the stresses on the lag screw In turn, this Indicates the existence of a location for the distal spew that compromises these two effects in an optimum way (C) 2009 Elsevier Ltd All rights reserved en
heal.publisher ELSEVIER SCI LTD en
heal.journalName INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED en
dc.identifier.doi 10.1016/j.injury.2009.09.006 en
dc.identifier.isi ISI:000276155600004 en
dc.identifier.volume 41 en
dc.identifier.issue 3 en
dc.identifier.spage 259 en
dc.identifier.epage 265 en


Αρχεία σε αυτό το τεκμήριο

Αρχεία Μέγεθος Μορφότυπο Προβολή

Δεν υπάρχουν αρχεία που σχετίζονται με αυτό το τεκμήριο.

Αυτό το τεκμήριο εμφανίζεται στην ακόλουθη συλλογή(ές)

Εμφάνιση απλής εγγραφής