MECHANICAL BLOOD TRAUMATIZATION BY TUBING AND THROTTLES IN INVITRO PUMP TESTS - EXPERIMENTAL RESULTS AND IMPLICATIONS FOR HEMOLYSIS THEORY

Αποθετήριο DSpace/Manakin

Εμφάνιση απλής εγγραφής

dc.contributor.author SCHIMA, H en
dc.contributor.author MULLER, MR en
dc.contributor.author TSANGARIS, S en
dc.contributor.author GHEISEDER, G en
dc.contributor.author SCHLUSCHE, C en
dc.contributor.author LOSERT, U en
dc.contributor.author THOMA, H en
dc.contributor.author WOLNER, E en
dc.date.accessioned 2014-03-01T01:42:12Z
dc.date.available 2014-03-01T01:42:12Z
dc.date.issued 1993 en
dc.identifier.issn 0160-564X en
dc.identifier.uri https://dspace.lib.ntua.gr/xmlui/handle/123456789/23714
dc.subject MECHANICAL HEMOLYSIS en
dc.subject SHEAR STRESS en
dc.subject BIOCOMPATIBILITY en
dc.subject CENTRIFUGAL BLOOD PUMP en
dc.subject.classification Engineering, Biomedical en
dc.subject.classification Transplantation en
dc.subject.other SHEAR-STRESS en
dc.title MECHANICAL BLOOD TRAUMATIZATION BY TUBING AND THROTTLES IN INVITRO PUMP TESTS - EXPERIMENTAL RESULTS AND IMPLICATIONS FOR HEMOLYSIS THEORY en
heal.type journalArticle en
heal.language English en
heal.publicationDate 1993 en
heal.abstract Blood has become essential as a test fluid to evaluate hemolysis and biocompatibility of blood pumps in vitro. The blood is usually pumped from a blood bag into a circuit against elevated pressure. A throttle or a length of tubing is used to produce the pressure head. Blood damage caused by the shear stress in these pressure-reducing devices should be minimal. It is not known whether the high but short-lasting shear stress in a throttle is more or less damaging to the blood than the low but long-lasting stress in tubing. In this study, throttles (width 11 mm, minimal height 0.9 mm, length 30 mm; shear stress = 136 N/m2 lasting for 3.23 ms); and tubing (inner diameter 9.5 mm, length 4.5 m, shear stress = 4.5 N/m2 lasting for 3.5 s) were compared at a flow of 5 L/min and a pressure drop of 150 mm Hg. Experiments (n = 10) with bovine blood were performed in two parallel setups using Bio-Medicus pumps BP80. Free hemoglobin in plasma (fHb) and thromboxane B2 (TXB2) were measured. After 6 h, the fHb increase was 31.9 +/- 19.1 mg% for the throttle setup and 32.3 +/- 16.2 for the tubing setup. The TXB2 release was 296 +/- 70 and 305 +/- 54 pg/0.1 ml respectively after 4 h. In summary, no significant differences between the two setups for either fHb or TXB2 could be detected. So the use of a throttle, which requires far less priming volume and a smaller blood-contacting surface while also offering a wider range of adjustment, seems preferable. Furthermore, in contrast to some other in vitro studies, these results strongly indicate that blood is quite insensitive to high shear stress exposure in the millisecond range even if repetitively applied. en
heal.publisher BLACKWELL SCIENCE INC en
heal.journalName ARTIFICIAL ORGANS en
dc.identifier.isi ISI:A1993KP61100003 en
dc.identifier.volume 17 en
dc.identifier.issue 3 en
dc.identifier.spage 164 en
dc.identifier.epage 170 en


Αρχεία σε αυτό το τεκμήριο

Αρχεία Μέγεθος Μορφότυπο Προβολή

Δεν υπάρχουν αρχεία που σχετίζονται με αυτό το τεκμήριο.

Αυτό το τεκμήριο εμφανίζεται στην ακόλουθη συλλογή(ές)

Εμφάνιση απλής εγγραφής