Thermoluminescent dosimetry of the selectSeed I-125 interstitial brachytherapy seed

Αποθετήριο DSpace/Manakin

Εμφάνιση απλής εγγραφής

dc.contributor.author Anagnostopoulos, G en
dc.contributor.author Baltas, D en
dc.contributor.author Karaiskos, P en
dc.contributor.author Sandilos, P en
dc.contributor.author Papagiannis, P en
dc.contributor.author Sakelliou, L en
dc.date.accessioned 2014-03-01T01:52:13Z
dc.date.available 2014-03-01T01:52:13Z
dc.date.issued 2002 en
dc.identifier.issn 0094-2405 en
dc.identifier.uri https://dspace.lib.ntua.gr/xmlui/handle/123456789/26603
dc.subject I-125 selectSeed en
dc.subject thermoluminescent dosimetry en
dc.subject Monte Carlo en
dc.subject brachytherapy en
dc.subject.classification Radiology, Nuclear Medicine & Medical Imaging en
dc.subject.other RADIATION-THERAPY COMMITTEE en
dc.subject.other MONTE-CARLO en
dc.subject.other WATER en
dc.subject.other RECOMMENDATIONS en
dc.subject.other ANISOTROPY en
dc.title Thermoluminescent dosimetry of the selectSeed I-125 interstitial brachytherapy seed en
heal.type journalArticle en
heal.language English en
heal.publicationDate 2002 en
heal.abstract This work presents experimental dosimetry results for the new selectSeed I-125 prostate seed design for use with the seedSelectron afterloading device, in accordance with the AAPM advisory that all new low energy interstitial brachytherapy seeds should undergo one Monte Carlo (MC) and at least one experimental dosimetry characterization. TLD dosimetry was per-formed using 120 cylindrical LiF TLD type-100 rods calibrated using a 6 MV photon beam. They were irradiated in solid water phantoms for the experimental determination of the seed dose rate constant, radial dose functions and anisotropy functions. MC simulations were performed for the determination of the TLDs relative energy response that was found position independent and equal to 1.40+/-0.03, and for the calculation of the ratio of dose in liquid water to dose in solid water that was found to be well described by D-liquid water/D-solid (water) = 1.013*r+0.030 presenting only a minor dependence on polar angle. The selectSeed dose rate constant in liquid water was found equal to 0.938+/-0.065 cGy h(-1) U-1, which agrees within experimental uncertainties with corresponding MC results of Lambda(select Seed) = 0.954+/-0.005 cGy h(-1) U-1. The experimental radial dose and anisotropy function results were also found in good agreement with corresponding MC calculations. (C) 2002 American Association of Physicists in Medicine. en
heal.publisher AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS en
heal.journalName MEDICAL PHYSICS en
dc.identifier.isi ISI:000175675000008 en
dc.identifier.volume 29 en
dc.identifier.issue 5 en
dc.identifier.spage 709 en
dc.identifier.epage 716 en


Αρχεία σε αυτό το τεκμήριο

Αρχεία Μέγεθος Μορφότυπο Προβολή

Δεν υπάρχουν αρχεία που σχετίζονται με αυτό το τεκμήριο.

Αυτό το τεκμήριο εμφανίζεται στην ακόλουθη συλλογή(ές)

Εμφάνιση απλής εγγραφής