HEAL DSpace

Performance of Hall sensor-based devices for magnetic field diagnosis at fusion reactors

Αποθετήριο DSpace/Manakin

Εμφάνιση απλής εγγραφής

dc.contributor.author Bolshakova, I en
dc.contributor.author Duran, I en
dc.contributor.author Holyaka, R en
dc.contributor.author Hristoforou, E en
dc.contributor.author Marusenkov, A en
dc.date.accessioned 2014-03-01T02:44:54Z
dc.date.available 2014-03-01T02:44:54Z
dc.date.issued 2007 en
dc.identifier.issn 1546-198X en
dc.identifier.uri https://dspace.lib.ntua.gr/xmlui/handle/123456789/32009
dc.subject Fusion reactor en
dc.subject Galvanomagnetic sensor en
dc.subject Magnetic diagnostics en
dc.subject Radiation hardness en
dc.subject.classification Chemistry, Analytical en
dc.subject.classification Electrochemistry en
dc.subject.classification Instruments & Instrumentation en
dc.subject.classification Physics, Applied en
dc.subject.other Actuators en
dc.subject.other Fusion reactors en
dc.subject.other Hall effect transducers en
dc.subject.other Magnetic field measurement en
dc.subject.other Radiation hardening en
dc.subject.other Galvanomagnetic sensor en
dc.subject.other Magnetic diagnostics en
dc.subject.other Radiation hardness en
dc.subject.other Magnetic sensors en
dc.title Performance of Hall sensor-based devices for magnetic field diagnosis at fusion reactors en
heal.type conferenceItem en
heal.identifier.primary 10.1166/sl.2007.004 en
heal.identifier.secondary http://dx.doi.org/10.1166/sl.2007.004 en
heal.language English en
heal.publicationDate 2007 en
heal.abstract In the paper, we report the results on the development of Hall sensor-based devices possessing the functions of selfdiagnostics and automatic correction, which are aimed at the improvement of the magnetic field measuring accuracy in radiation environment of thermonuclear reactors. Developed devices are based on radiation-hard semiconductor sensors. The method for device stabilization is offered, which is based on a test method of measurements, where sensor and actuator are combined in a single primary transducer. New algorithm of correction of the transduction function on the basis of frequency separation of differential and integral signal components is proposed. The results of the testing of developed Hall sensor-based devices, applied for the magnetic field measurement at the largest European thermonuclear reactors TORE SUPRA (France) and JET (Great Britain), are presented. Copyright © 2007 American Scientific Publishers. All rights reserved. en
heal.publisher AMER SCIENTIFIC PUBLISHERS en
heal.journalName Sensor Letters en
dc.identifier.doi 10.1166/sl.2007.004 en
dc.identifier.isi ISI:000245571600074 en
dc.identifier.volume 5 en
dc.identifier.issue 1 en
dc.identifier.spage 283 en
dc.identifier.epage 288 en