HEAL DSpace

432-MHz LOCAL HYPERTHERMIA SYSTEM USING AN INDIRECTLY COOLED, WATER-LOADED WAVEGUIDE APPLICATOR.

Αποθετήριο DSpace/Manakin

Εμφάνιση απλής εγγραφής

dc.contributor.author Uzunoglu Nikolaos, K en
dc.contributor.author Angelikas Elias, A en
dc.contributor.author Cosmidis, PA en
dc.date.accessioned 2014-03-01T01:06:43Z
dc.date.available 2014-03-01T01:06:43Z
dc.date.issued 1987 en
dc.identifier.issn 0018-9480 en
dc.identifier.uri https://dspace.lib.ntua.gr/xmlui/handle/123456789/9583
dc.subject.classification Engineering, Electrical & Electronic en
dc.subject.other BIOMEDICAL ENGINEERING - Hyperthermia Therapy en
dc.subject.other WAVEGUIDES - Microwaves en
dc.subject.other INDIRECT COOLING en
dc.subject.other LOCAL HYPERTHERMIA SYSTEM en
dc.subject.other WAVEGUIDE ANTENNA en
dc.subject.other ANTENNAS en
dc.title 432-MHz LOCAL HYPERTHERMIA SYSTEM USING AN INDIRECTLY COOLED, WATER-LOADED WAVEGUIDE APPLICATOR. en
heal.type journalArticle en
heal.identifier.primary 10.1109/TMTT.1987.1133611 en
heal.identifier.secondary http://dx.doi.org/10.1109/TMTT.1987.1133611 en
heal.language English en
heal.publicationDate 1987 en
heal.abstract A water-loaded indirectly cooled waveguide antenna with a small aperture (2 multiplied by 5. 5 cm**2) has been developed for local hyperthermia applications. The operation frequency is 432 MHz. The indirect cooling is achieved by circulating water in an external chamber having common walls with the waveguide, and good thermal and electrical stability is achieved for radiation powers up to 100 W. The water loading provides a good coupling of the radiated power into tissue. The observed variations in standing wave ratios and the concentration of the specific absorption rates (SARs) are found to be satisfactory for localized superficial hyperthermia treatments. The matching of the water-loaded antenna into the power generator requires special attention because of two possible matching conditions corresponding to high- and low-power dissipations inside the antenna cavity. SAR distributions inside tissues have been computed and compared with thermographic measurements performed in phantoms. Preliminary results of trials in Wistar rats transplanted with Walker sarcoma are also reported. en
heal.publisher IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC en
heal.journalName IEEE Transactions on Microwave Theory and Techniques en
dc.identifier.doi 10.1109/TMTT.1987.1133611 en
dc.identifier.isi ISI:A1987F767000004 en
dc.identifier.volume MTT-35 en
dc.identifier.issue 2 en
dc.identifier.spage 106 en
dc.identifier.epage 111 en